Western Blot(WB)

Issue Type

Core Causes

Solutions

No Target Band

1. Primary/secondary antibody inactivation or improper species/antigen matching;

 

2. Incorrect electrophoresis polarity or membrane transfer orientation;

 

3. Excessive blocking or over-washing;

 

4. Expired luminescent substrate

1. Verify antibody specificity and expiration, store in aliquots to avoid repeated freeze-thaw;

 

2. Confirm correct assembly of electrophoresis and transfer components;

 

3. Optimize blocking (1-2h at RT with 5% non-fat milk/3% BSA) and washing (3×5-10min with TBST);

 

4. Use fresh, light-protected substrate

Weak Band Signal

1. Low protein concentration or incomplete extraction;

 

2. Insufficient transfer time/voltage;

 

3. Over-diluted antibodies;

 

 4. Inadequate incubation/exposure time

1. Increase sample amount or optimize extraction with inhibitors;

 

2. Adjust transfer parameters by protein molecular weight;

 

3. Follow antibody dilution recommendations (primary: 1:500-1:10000, secondary: 1:10000-1:50000);

 

4. Extend incubation (primary: 4℃ overnight) or exposure (1-5min)

High Background (Overall Membrane Glow)

1. Excess antibody concentration;

 

2. Inadequate blocking or incomplete washing;

 

3. Membrane drying during incubation/washing;

 

4. Excessive substrate or prolonged incubation

1. Reduce antibody concentration;

 

2. Extend blocking time (2h at RT) and increase washing frequency (4-5×10min with TBST);

 

3. Ensure membrane remains fully submerged;

 

4. Blot excess substrate and shorten incubation to 1-2min

Multiple Non-Specific Bands

1. Poor primary antibody specificity;

 

2. Protein degradation or cross-reactivity of secondary antibody;

 

3. Insufficient blocking/washing

1. Switch to monoclonal antibodies;

 

3. Add protease inhibitors during extraction and use species-matched secondary antibody;

 

3. Optimize blocking buffer (3% BSA for phosphorylated antigens) and washing conditions

Blurred/Tailed Bands

1. Protein aggregation/degradation;

 

2. High sample salt concentration or excessive loading;

 

3. Improper gel concentration or electrophoresis voltage

1. Avoid repeated freeze-thaw and add fresh reducing agents;

 

3. Desalt samples or reduce loading volume;

 

3. Select gel concentration based on protein size (8-12% for 30-100kDa) and use constant voltage (80V stacking gel, 120V resolving gel)

Abnormal Transfer

1. Insufficient/excessive transfer time/voltage;

 

2. Bubble formation in transfer sandwich;

 

3. Incorrect membrane pore size;

 

4. Abnormal methanol concentration in transfer buffer

1. Adjust transfer parameters (20-30min for <30kDa, 60-90min for >100kDa);

 

2. Remove bubbles during sandwich assembly;

 

3. Choose 0.2μm membrane for <20kDa, 0.45μm for others;

 

4. Maintain 20% methanol in transfer buffer

Uneven Bands (Spotted/Striped)

1. Uneven antibody distribution (insufficient shaking);

 

2. Precipitates in reagents;

 

3. Inadequate reagent volume covering the membra

1. Incubate on a shaker (50-100rpm);

 

3. Centrifuge reagents (10000rpm, 5min) to remove precipitates;

 

3. Ensure sufficient reagent volume (≥100μL/cm² membrane)


Related Promotional Journal Downloads

Brand/Complete Product Range Series
Brand/Complete Product Range Series

Detailed introduction of the column

Download and Save
Primary antibodies
Primary antibodies

Detailed introduction of the column

Download and Save
Secondary antibodies
Secondary antibodies

Detailed introduction of the column

Download and Save
mIHC/ Multiplex Fluorescence Detection Kits
mIHC/ Multiplex Fluorescence Detection Kits

Detailed introduction of the column

Download and Save

Explore Our Recommended Popular Products

More products

30,000+ high- quality products available online

Primary Antibodies, Secondary Antibodies, mIHC Kits, ELISA Kits, Proteins, Molecular Biology Products,Cell Lines,Reagents ...

Contact Us