Issue Type

Phenomenon Description

Core Causes

Solutions

No/Low Fluorescent Signal

No fluorescence in target cell population or signal close to negative control

1. Insufficient cell permeabilization (for intracellular antigen detection);

 

2. Fluorochrome quenching or low labeling efficiency;

 

3. Low cell concentration or antigen expression;

 

4. Antibody inactivation, mismatched species or over-dilution;

1. Permeabilize cells with 0.1% Triton X-100 for 10–15min for intracellular staining, omit for membrane protein detection;

 

2. Operate in dark throughout and use high-activity fluorochromes;

 

3. Concentrate cells to 1×10⁶–1×10⁷ cells/mL or use signal amplification reagents;

 

4. Verify antibody validity, optimize dilution ratio (1:50–1:200) and match target cell species;

High Non-specific Background

Elevated fluorescence in negative cell population, unclear population separation

1. Excessive antibody concentration;

 

2. Inadequate blocking (Fc receptor interference);

 

3. Excessive cell debris/dead cells causing non-specific antibody adsorption;

 

4. Incomplete washing with residual unbound antibody

1. Reduce antibody concentration

 

2. Block with 1% BSA or species-matched serum for 15–20min, add Fc receptor blocker for Fc-positive cells;

 

3. Filter cells through nylon mesh before assay, exclude dead cells with trypan blue/7-AAD;

 

4. Wash with PBS containing 2% FBS, discard supernatant completely after centrifugation and repeat 3 times

Cell Aggregation

Massive clump signals in scatter plot, unable to gate accurately

1. Over/under-digestion of adherent cells;

 

2. Insufficient pipetting for cell dispersion;

 

3. Excessively high centrifugation speed causing cell compression;

 

4. Inadequate mixing during antibody incubation

1. Optimize digestion conditions (e.g., trypsinize until cell edge retraction), avoid over-digestion;

 

2. Pipette cells gently with Pasteur pipette or pass through 200-mesh sieve;

 

3. Reduce centrifugation speed (300–400×g for 5min);

 

4. Mix gently by vortex or incubate on shaker at low speed

Low Resolution, Poor Population Separation

Overlap between target and miscellaneous populations, unable to distinguish

1. Improper fluorescence compensation setting;

 

2. Low laser power reducing detection sensitivity;

 

3. Overlapping emission spectra of antibody fluorochromes;

 

4. Poor cell culture status with uneven antigen expression

1. Adjust fluorescence compensation with single-stained tubes to eliminate inter-channel interference;

 

2. Increase laser power appropriately (avoid excessive cell damage);

 

3. Select fluorochrome combinations with non-overlapping emission spectra (e.g., FITC/PE/Cy5);

 

4. Use cells in logarithmic growth phase and standardize culture conditions

Low Cell Viability

Dead cell ratio > 20% during detection

1. Prolonged sample processing time causing hypoxia-induced cell damage;

 

2. Severe centrifugation/pipetting leading to mechanical damage;

 

3. Improper osmotic pressure of staining solution causing cell rupture;

 

4. Toxic effects of antibodies/reagents

1. Shorten sample processing time (<2h) and operate on ice throughout;

 

2. Pipette gently and keep centrifugation speed below 500×g;

 

3. Prepare staining solution with isotonic PBS and avoid high-concentration reagents;

 

4. Select low-toxicity antibodies and optimize incubation time (15–30min)

Poor Experimental Reproducibility

Large variations in results between batches and obvious fluctuation in cell proportion

1. Inconsistent staining temperature/time;

 

2. Unstable flow rate causing data acquisition errors;

 

3. Different initial cell status (passage/density variations)

1. Fix staining conditions (incubate at 4℃ for 30min in dark);

 

2. Calibrate instrument before detection and maintain stable flow rate (100–300 events/sec);

 

3. Use cells of the same passage and in logarithmic growth phase


Related Promotional Journal Downloads

Brand/Complete Product Range Series
Brand/Complete Product Range Series

Detailed introduction of the column

Download and Save
Primary antibodies
Primary antibodies

Detailed introduction of the column

Download and Save
Secondary antibodies
Secondary antibodies

Detailed introduction of the column

Download and Save
mIHC/ Multiplex Fluorescence Detection Kits
mIHC/ Multiplex Fluorescence Detection Kits

Detailed introduction of the column

Download and Save

Explore Our Recommended Popular Products

More products

30,000+ high- quality products available online

Primary Antibodies, Secondary Antibodies, mIHC Kits, ELISA Kits, Proteins, Molecular Biology Products,Cell Lines,Reagents ...

Contact Us